Advanced ImagingRP-CT1 podNewPreliminaryLicensed service

Imaging CT

40-row CT for Port Health diagnostics, in one pod.

Imaging CT floor plan, RP-CTSCHEMATIC · PRELIMINARYREINFORCED FLOORGANTRYTABLECONSOLECABINETSCORRIDORPATIENTCHANGE44 SFSCAN ROOM165 SFCONTROL55 SFEQUIPMENT55 SFCORRIDOR5'-0"4'-0"15'-0"5'-0"5'-0"5'-0"40'-0" exterior12'-0" ext.

Schematic layout (preliminary), drawn on the 12'-0" × 40'-0" pod (11'-0" × 39'-0" interior) · ~430 sq ft · dashed walls mark where the next pod connects

What it adds

  • Scan room for a United Imaging uCT 528, larger than the 9 m² (approx. 97 SF) minimum installation area
  • Control room beside the scan room, with a shielded viewing window
  • Equipment area for the CT cabinets and power distribution
  • Patient change area opening into the scan room

Building systems

  • Radiation shielding
  • Dedicated power
  • HVAC heat load
  • Reinforced floor

Rooms

Patient Change44 SF
Scan Room165 SF
Control55 SF
Equipment55 SF

Furniture & equipment

  • • United Imaging uCT 528 CT scanner (40-row)
  • • Patient table
  • • CT control console and workstation
  • • Shielded viewing window
  • • Operator chair
  • • CT equipment cabinets
  • • Power distribution unit
  • • Change bench and lockers
  • • Lead apron rack

Preliminary. Confirm with vendor site-planning guide.

Fit check

Pods required1 pod
Floor area~430 SF
Point loadCT gantry, 1,800–3,500 kg
PowerDedicated supply; HVAC heat load
ShieldingLead, calculated by a physicist

Preliminary. Confirm with vendor site-planning guide.

Equipment values

  • UnitPreliminary

    United Imaging uCT 528 (40-row CT)

  • Minimum installation areaPreliminary

    9 m² (approx. 97 SF)

  • Gantry weightPreliminary

    1,800 to 3,500 kg (typical CT), taken as a concentrated point load

Every value above: Preliminary. Confirm with vendor site-planning guide.

Design notes

  • Lead shielding: a qualified physicist calculates the thickness and extent for the site and workload
  • Dedicated power supply for the CT
  • HVAC sized for the CT and equipment heat load
  • Floor reinforcement under the gantry for the concentrated point load